Retsch ZM 200 Bedienungsanleitung Seite 27

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 32
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 26
08.02.2012 Retsch GmbH 27 0003
Error messages on the display
F03 to F12
The display shows
The display shows
F03
F04
PROBLEM WITH
CLOSE OR OPEN COVER,
THE INTERLOCK
SHOULD PROBLEM REMAIN
THEN INTERLOCK FAULTY
SERVICE REQUIRED!
SERVICE REQUIRED!
The error appears when the cover
is closed and the interlock is
faulty.
Turn the device off at the
main switch and then
turn on again.
If the error reappears, there is a
safety problem and servicing is
required.
The error appears when the cover
is closed and the interlock is
faulty.
Turn the device off at the
main switch and then
turn on again.
If the error reappears, there is a
safety problem and servicing is
required.
F05
F06
The display shows
The display shows
CLOSE COVER
OPEN COVER
The display shows
F07
FAULTY MOTOR
CONTROL
SERVICE REQUIRED!
The display shows
The display shows
F09
F10
COOLING FAN IS OFF
ROTOR DOES NOT ROTATE
BLOCKED ?
CASSETTE COVER
SERVICE REQUIRED!
MISSING ?
The error appears if the rotor is
unable to turn. Possible causes
may be a missing cassette cover,
material between the rotor and
ring sieve or the motor is
defective.
F12
The display shows
YOUR SERVICE INTERVAL
HAS EXPIRED
PLEASE SERVICE UNIT
CONTINUE WITH 'STOP'
The error appears when the
service interval programmed by
the customer has expired.
Seitenansicht 26